A chromatogram displays a signal as material passes through a separation system. A peak table summarizes features the lab assigned within that trace. Read both alongside the method and sample identifiers to see how the laboratory got from the displayed signal to the reported result.
Start with the axes and trace identifiers
The horizontal axis generally shows time or elution volume. The vertical axis shows detector response or another reported measure of concentration. Check the units and scale on each axis, and look for a sample or run identifier that links the plot to the certificate. A cropped image showing only the tallest peak provides a more limited view than the complete report.
If you are comparing separate plots, confirm that the scales, detection channels and method versions are comparable. Changing the vertical scale can make a secondary peak appear more or less prominent even when the measured signals are similar.
Read the peak table alongside the plot
Retention time indicates where the laboratory recorded the signal. Integrated area estimates the response across the assigned peak. If an area percentage is reported, find the denominator and any peaks excluded by the laboratory’s rules. A peak that accounts for most of the integrated signal under one method still needs an appropriate response model before that area is interpreted as a mass fraction.
Look for peaks that appear close together, prominent shoulders and any integration notes. The trace alone can suggest that reviewers should ask about resolution. Adequate separation and the handling of overlapping responses need evidence from the laboratory’s validated method and run records.
Ask for the link from the plot to the stated claim
A report that lists a target peak’s retention time should explain the basis for assigning it. A quantitative claim should identify the calculation or reporting basis that translates the relevant response to a value. The amount of documentation needed depends on what the certificate claims.
Useful follow-up requests include the sample-linked chromatogram, full peak table, method identifier and any notes on manual integration or omitted peaks. For an identity or mass claim, ask which additional measurement ties the plotted result to the specific property being certified.