A report can show 99% HPLC purity and a measured amount below the label claim. The two figures can both be accurate because they describe different measurements.
What HPLC purity measures
When purity is calculated by peak area, 99% means that roughly 99% of the integrated area was assigned to the target peak under that method. The result depends on what the method separates and detects. Water and counterions can require separate measurements.
Where to find the amount per vial
Look for a result labeled content, assay, or quantity, then check its units and reporting basis. In a hypothetical report for a vial labeled 10 mg, a purity result of 99% and a content result of 8.7 mg per vial mean the lab measured 8.7 mg per vial. Multiplying the nominal 10 mg by the purity percentage would give a different figure.
For background on chromatograms, see HPLC Analysis Explained and the USP peptide quality paper.