A reorder can arrive with the same product number but a new lot. The practical question is whether the records support using the new material under the project’s existing quality criteria.
Save the before and after
Pull together the previous decision, certificate versions for both lots, and any change notice from the supplier. Compare the product description, assay basis, methods, acceptance criteria, and reporting units. A difference in the report template may be cosmetic. A change in basis or scope affects whether two reported numbers can be compared directly.
Use the study’s quality criteria
Different projects may respond differently to the same change. A lab comparing observations across several months might need explicit evidence of comparability. A separate short study may have a simpler acceptance decision. Record the criteria originally approved for the project and evaluate the replacement against those criteria.
Imagine the old certificate provides a numeric result and the new one says only that it passes. Both can be valid release documents, but the new certificate may give you less evidence for a trend comparison. Ask whether the issuer can provide a report with the comparable result and method version.
Write the next decision, including any limits
Record the person who assessed the change, the date, affected projects, supporting documents, and any pending evidence. If older work is unaffected, document why. If a supplier cannot tell you when the change took effect, keep the open question separate from the records you have already checked.